Flex TENS ™ Protocol


 Click on the diagnosis below for suggested electrode placements
                                                                     
                                                                              

TARSAL TUNNEL SURGERY

 

Flex TENS ™ Tarsal Tunnel Surgery protocol diagram

Electrode Placement

Along medial and lateral border of foot

Control Settings

Mode: 8/8 or 10/10
Frequency: 100
Treatment Time: 15 minutes
Amplitude: Maximum Sensory Level

THEN

Mode: 1/1
Frequency: 10
Treatment Time: 15 minutes
Amplitude: Maximum Sensory Level

FLEX-GAR SOCK™

  • Red lead goes into sock
  • Black lead placed in electrode proximal to the body
 


LOW BACK PAIN

 

Flex TENS ™ Low Back Pain protocol diagram

Electrode Placement

OPTION 1: Centralized LBP
Channel 1: Bilaterally at level L2-3
Channel 2: Bilaterally at SI level

Note: General rule - place channel
1 above pain level, channel 2 below
pain level

OPTION 2: Radiating LBP
Channel 1: Bilaterally at lumbar level involved
Channel 2: One patch over SI joint of affected side, other over sciatic notch or just distal to end point of pain

Control Settings

Mode: N
Frequency: 125 pps
Pulse Width: 20 - 60 ms
Amplitude: Sensory level to tolerance
Treatment Time: 30 minutes to 1 hour intermittently throughout the day

FLEX-GAR BACK™

  • Place electrodes as shown in option 1


CERVICAL SPRAIN/STRAIN

 

Flex TENS ™ Low Back Pain protocol diagram

Electrode Placement

Channel 1: Place electrodes bilaterally at suboccepital area
Channel 2: Bilaterally at C7-T1 level

Control Settings

Mode: N
Frequency: 10 pps
Pulse Width: 150 - 250 ms
Amplitude: Sensory level to tolerance
Treatment Time: 30 minutes to 1 hour intermittently throughout the day

FLEX-GAR CERVICAL™

  • Place electrodes as shown
 


SHOULDER SPRAIN

 

Flex TENS ™ Shoulder Sprain protocol diagram

Electrode Placement

OPTION 1: Centralized LBP
Channel 1: Place one pad on trigger point of supraspinatus, other on deltoid insertion
Channel 2: One pad on pec insertion other one teres major

Control Settings

Mode: N
Frequency: 100 pps
Pulse Width: 150 ms
Amplitude: Turn up to MS contraction state if palpable trigger points
Treatment Time: 15 - 20 minutes to decrease trigger points

FLEX-GAR SHOULDER™

  • Place electrodes as shown
 


HIP STRAIN/PAIN

 

Flex TENS ™ Shoulder Sprain protocol diagram

Electrode Placement

  • May use 1, 2, or 4 electrodes
  • If you use 1 pad try paravertebral
  • If you use 2 pads use 2" x 3.5"
  • If you use 4 pads use 2" round

OPTION 1
Place 2" round pads 3 - 4 inches
apart forming a square around the site of pain

OPTION 2
Place 2" x 3.5" above and below site of pain

OPTION 3
Paravertebral Pad - Place over involved region

Control Settings

Mode: N
Frequency: 80 pps
Pulse Width: 250 ms
Amplitude: Just below MS contraction

Alternate Control Settings

Mode: N
Rate: 4 pps
Duration: 150 ms
Amplitude: Just below MS contraction
Treatment Time: 15 - 20 minutes


FOOT OR ANKLE PAIN

 

Flex TENS ™ Shoulder Sprain protocol diagram

Control Settings

CHRONIC
Mode: M
Frequency: 150
Pulse Width: 70 ms
Amplitude: Maximum Sensory Level
Treatment Time: 20 - 30 minutes

Interferential

CHRONIC
Mode: 10/10
Frequency: 80 - 150
Treatment Time: 20 - 30 minutes
Amplitude: Maximum Sensory Level

EDEMA
Mode: 1/1
Frequency: 1 - 10
Treatment Time: 20 - 30 minutes
Amplitude: Maximum Sensory Level

back to top