IF-5000 ™ Combo Protocol


 
 Click on the diagnosis below for suggested electrode placements:
 
                                                                         

ACL RECONSTRUCTION

IF 5000™ ACL RECONSTRUCTION protocol diagram

Goal: lower edema, facilitate quad set.

Electrode Placement

A - Channel A (IFC) - Quadripolar placement around knee for IFC

B - Channel B (NMS) - One Pad on VMO, one pad on rectus femoris

Control Settings

IF - LOW
CB - IF - 15 minutes
MS - W - 400
CB - MS - 15 minutes
Treatment Time: 30 minutes

TO LOCK UNIT: Press ENTER once and press and hold again for 5 seconds (unit will display "LCKD MS").

 


FIBROMYALGIA

 

IF 5000™ FIBROMYALGIA protocol diagram

Electrode Placement

Find trigger points along paraspinals

Control Settings

A & B - IF-WIDE
Treatment Time: 20 minutes

FLEX-GAR™ THORACIC

  • Place electrodes as shown

 

 

back to top


CHRONIC LOWER BACK PAIN & ACUTE CERVICAL PAIN

 

IF 5000™ CHRONIC LOWER BACK PAIN & ACUTE CERVICAL PAIN protocol diagram

Control Settings

A - Chronic Lower Back Pain    IF - HIGH
   Treatment Time: 20 - 30 minutes

B - Acute Cervical Pain    IF - WIDE
   Treatment Time: 20 - 30 minutes

FLEX-GAR™ LOWER BACK
FLEX-GAR™ CERVICAL

  • Place electrodes as shown

 

 

back to top


SHOULDER PAIN

 

IF 5000™ SHOULDER PAIN protocol diagram

Non Surgical

Control Settings

IF - High
Treatment Time: 15 minutes

Surgical

Control Settings

IF - Low
Treatment Time: 15 minutes

 

 

 

back to top


EPICONDYLITIS

 

IF 5000™ EPICONDYLITIS protocol diagram

Lateral

Control Settings

IF - HIGH
Treatment Time: 15 minutes

Medial

Control Settings

Same Settings

 

 

 

 

 back to top


KNEE PAIN

 

IF 5000™ KNEE PAIN protocol diagram

Control Settings

EDEMA REDUCTION
IF - LOW
Treatment Time: 15 - 20 minutes

CHRONIC PAIN
IF - HIGH
Treatment Time: 15 - 20 minutes

ACUTE PAIN
IF - WIDE
Treatment Time: 15 - 20 minutes

 

 

 

 back to top


PLANTAR FASCIITIS

 

IF 5000™ PLANTAR FASCIITIS protocol diagram

Electrode Placement

Electrodes are placed on medial & lateral side of foot

Control Settings

IF - WIDE
Treatment Time: 10 minutes

THEN

IF - LOW
Treatment Time: 10 minutes

 

 

back to top


BONE/SOFT TISSUE INJURY OF FOREFOOT

 

IF 5000™ BONE/SOFT TISSUE INJURY OF FOREFOOT protocol diagram


Electrode Placement

Electrodes are placed on dorsal and plantar surface in tender area

Control Settings

IF - Low
Treatment Time: 15 minutes

 

 

 

back to top


 

WRIST STRENGTHENING

 

IF 5000™ WRIST STRENGTHENING protocol diagram

Electrode Placement

Channel 1:
Electrode over proximal third of forearm over wrist extensor bellies
Electrode ~ 2" superior to wrist crease in center of forearm

Channel 2:
Electrode placed over proximal third of forearm over wrist flexor bellies
Electrode placed centrally over flexor tendon ~ 3" superior to wrist crease

Perform only 1 channel at a time or alternate setting when using NMES.

Control Settings

MS - W - 300
MS - CTR - 08
MS - RLx - 02
Amplitude: MS contraction
Treatment Time: 10 minutes

 back to top


INTRINSIC STRENGTHENING

 

IF 5000™ INTRINSIC STRENGTHENING protocol diagram

Electrode Placement

(-) Just proximal to pisiform
(+) 2 - 3" superior and lateral to (-) electrode

Control Settings

MS - W - 300
MS - CTR - 10
MS - RLx - 30
MS - RMP - 03
Treatment Time: 15 minutes

 

 

back to top


GRIP STRENGTHENING

 

IF 5000™ GRIP STRENGTHENING protocol diagram

Recommend dual channel to stimulate intrinsics and wrist flexors

Electrode Placement

Use 2" round electrodes and place 2 - 3 inches superior to volar crease. Place (+) leads from both channels here.

Place channel 1 (-) electrode proximal to pisiform.

Place channel 2 (-) electrode 1 - 2 inches distal to medial epicondyle over flexor MS bellies.

Control Settings

MS - W - 350
MS - CTR - 10 seconds
MS - RLx - 35 seconds
MS - RMP - 02
Treatment Time: 15 - 20 minutes

 

back to top


ANTERIOR SHOULDER SUBLUXATION

 

IF 5000™ ANTERIOR SHOULDER SUBLUXATION protocol diagram

Electrode Placement

Place one electrode over posterior deltoid, other over supraspinatus

Control Settings

MS - W - 350
MS - CTR - 15 seconds
MS - RLx - 3 seconds
MS - RMP - 03
Treatment Time: 15 - 20 minutes

 

 

 

back to top


ANTERIOR TIBIALIS STRENGTHENING

 

IF 5000™ ANTERIOR TIBIALIS STRENGTHENING protocol diagram

Electrode Placement

Place one electrode on superior lateral MS belly of tibialis anterior, other approx. 3 - 4" below staying on muscle bulk of anterior tib.

Control Settings

MS - W - 350
MS - CTR - 10 seconds
MS - RLx - 20 seconds
MS - RMP - 02
Treatment Time: 15 - 20 minutes

 

 

 

 back to top

Supply Contact Form.

Your Name
Date of Birth
Phone Number
(with Area Code, numbers only)